凝聚核粒子计数器原理与实验研究
DOI:
CSTR:
作者:
作者单位:

作者简介:

通讯作者:

中图分类号:

TH724

基金项目:

国家自然科学基金(52065002)、宁夏自然科学基金(2022AAC02045)项目资助


Principle and experimental study of the condensed particle counter
Author:
Affiliation:

Fund Project:

  • 摘要
  • |
  • 图/表
  • |
  • 访问统计
  • |
  • 参考文献
  • |
  • 相似文献
  • |
  • 引证文献
  • |
  • 资源附件
  • |
  • 文章评论
    摘要:

    针对环境中纳米量级颗粒物浓度检测问题,对凝聚核粒子计数器进行了研究与设计。首先对凝聚核粒子生长原理及 检测原理进行了研究,并依据异质核化原理设计了凝聚核粒子计数器的生长装置;其次设计了传感器后级信号处理电路用于 解决微弱粒子检测信号抗干扰能力差和测量困难问题;然后设计了基于抗饱和积分PID的控制系统,实现了采样流量的闭环 控;最后制研制了凝聚核粒子计数器样机,并开展了流量控制、纳米量级粒子生长及计数实验。实验结果表明,设计的凝聚核 粒子计数器,流量稳定性在2%以内,以TSI3772型凝聚核粒子计数器为标准仪器的计数相对误差在士8.5%以内,满足了行 业规范要求。

    Abstract:

    Aiming at the problem of nano-scale particle concentration detection in the environment,the condensed particle counter was researched and designed.Firstly,the growth principle and detection principle of condensed particles were researched,and the growth device of condensed particle counter was designed according to the heteronuclear principle. Secondly,the signal processing circuit of the sensor was designed to solve the problem of poor anti-interference ability and difficult measurement of weak signal of particle detection.Then,a control system based on anti-saturation integral PID was designed to realize the closed-loop control of sampling flow.Finally,a prototype of condensed particle counter was developed,and several experiments were carried out,including flow control,nano-scale particle growth and counting.The experimental results show that the flow stability of the condensation particle counter designed is within 2%,and the relative error of counting with the TSI3772 condensation particle counter as the standard instrument is within±8.5%,which meets the requirements of the industry standard.

    参考文献
    相似文献
    引证文献
引用本文

张白,蒋尊阳,曹 威,孔德超.凝聚核粒子计数器原理与实验研究[J].国外电子测量技术,2024,43(6):94-101

复制
分享
文章指标
  • 点击次数:
  • 下载次数:
  • HTML阅读次数:
  • 引用次数:
历史
  • 收稿日期:
  • 最后修改日期:
  • 录用日期:
  • 在线发布日期: 2024-07-09
  • 出版日期:
文章二维码
×
《国外电子测量技术》
财务封账不开票通知