Abstract:This paper describes a lowvoltage ACDC transfer traceability technology at 2-500 mV/1 kHz-1 MHz range, it also describes the use of a decimal inductive voltage divider used for the scale expansion of ACDC transfer technology. Thompson method is introduced in the paper and it gives the experimental data of calibration of IVD. This article also describes the design and winding of a calibration IVD with an additional secondary winding. Correction circuit and inputoutput isolation voltage divider, and Ntype connector is designed to further reduce the magnetic error. The calibration system is applied at millivolt voltage and industrial frequencies.